hematology machine 
Performance
Back spectro photometry, 8 hard membrane filter, separate signal process circuit for each wavelength, filters
select completed electronically, no mechanical movement ;
Hermetic solid state optical system, low contamination of dust and dampness, long life time of hard membrane filters;
Titanium alloy flow cell with small capacity and minimize cross-contamination;
1000 analysis data can be saved;
Clock CMOS chip inserted ;
256x65 dot-matrix back light LCD, display character curve simultaneously;
Reasonable software, programmable biochemical test menu, 200 test program can be saved;
Self-test program and trouble prompt are avaiable , test error<1%, CV<3% , cross-contamination<1%;
Technical feature
Analysis method: kinetics, end point, two-point, etc ;
Power supply:20w/12v halogen light, life time:3000hr;
Optical system: 8 hard membrane filter: 340, 380, 405, 492, 510, 546,578,630nm(wave length) Accuracy ±2nm;
ABS range:0-2.500Abs, resolution:0.001Abs;
Flow cell: Titanium quartz alloy, 33ul, optical path:10mm ;
Temperature: PELTIER elements, 25℃, 30℃,37℃±0.1℃ ;
Precision: error<1%,drift<5mAbs/hr ;
Printer: 58mm thermal printer;
Keyboard: 24 keys touch keyboard;
Power supply: 220V±15%, 50Hz, 70VA;
Interface: standard RS232 serial interface;
Dimensions: 460(L)X380(W)X200(H)mm;
Weight: 15kg;
Condition
Temperature: 0℃-40℃;
Humidty:5-80%;